• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE Back to Search Results
Model Number 2800TFX
Medical Device Problem Codes Calcified (1077); Degraded (1153); Fluid/Blood Leak (1250); Gradient Increase (1270); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Dyspnea (1816); Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Date of Event 02/09/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
Through implant patient registry it was learned that a patient with a 25mm 2800tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 5 years, 2 months due to unknown reason.The tavr was performed with a 26mm 9750tfx transcatheter valve.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional "customer complaint".The information reported may or may not be related to the edwards device.
 
Event or Problem Description
Through implant patient registry it was learned that a patient with a 25mm 2800tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 5 years, 2 months due to severe stenosis with insufficiency.The patient presented with shortness of breath.The tavr was performed with a 26mm 9750tfx transcatheter valve.The patient was in stable condition post procedure.
 
Additional Manufacturer Narrative
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.The device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.D8 - "no" should have been selected on initial fda report.
 
Event or Problem Description
Through implant patient registry and medical records, it was learned that a patient with a 25mm 2800tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 5 years, 2 months due to leaflets calcification and degeneration.The patient presented with heart failure.The tavr was performed with a 26mm 9750tfx transcatheter valve.The patient was in stable condition post procedure.
 
Additional Manufacturer Narrative
Calcific degeneration is a common cause of bioprosthetic heart valve failures.Many factors contribute to the onset and propagation of calcification.These include patient factors (age, disease state, pharmacological intervention, etc.), mechanical stress related to the valve's hemodynamic performance, and glutaraldehyde fixation of tissue.Of these, the fixation process is a relatively minor contributor to calcification for edwards' tissue valves due to anti-calcification treatments during manufacturing.Though numerous studies have been conducted on preventive calcification strategies in bioprosthetic heart valves, the causes of calcification are not fully understood and there are still no mechanisms or medical therapies which fully prevent bioprostheses from calcifying.Calcification is most commonly related to patient factors and is not usually an indication of a device malfunction.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key16694060
Report Number2015691-2023-12146
Device Sequence Number6080721
Product Code DYE
UDI-Device Identifier00690103194104
UDI-Public(01)00690103194104(17)201207
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/07/2020
Device Model Number2800TFX
Device Catalogue Number2800TFX25MM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2023
Supplement Date Received by Manufacturer04/12/2023
05/10/2023
Initial Report FDA Received Date04/06/2023
Supplement Report FDA Received Date05/08/2023
05/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/08/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexMale
-
-