|
STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 36/+7,5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
|
Back to Search Results |
|
| Model Number |
6570-0-736 |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Codes |
Memory Loss/Impairment (1958); Pain (1994); Metal Related Pathology (4530)
|
| Date of Event |
06/06/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Rep reported the patient's left hip implanted on (b)(6) 2011 was revised on (b)(6) 2023.Initial patient complaints were pain, elevated ion levels, and memory loss.Intra-operatively, corrosion was noted at the stem/ neck interface and at the interface of the trunnion and ceramic head.The stem construct, head, and liner were revised.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding corrosion involving a ceramic head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the event could not be confirmed as insufficient information was provided.Further information such as device return, pre- and post-operative x-rays, primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Search Alerts/Recalls
|
|
|