| Model Number |
FINETUNER ECHO [FINETUNER 2] |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
A finetuner echo was returned to service and repair.Upon preliminary investigation at service and repair it was determined there is failure of the tantalum capacitor.The stated reason for return to service and repair was due to non-function.To date, no injuries have been reported.
|
| |
|
Event or Problem Description
|
|
A finetuner echo was returned to service and repair.The stated reason for return to service and repair was due to non-function.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: according to the information received, a finetuner echo was sent to service repair due to not functioning.During device investigation a failed capacitor was detected which was clearly the reason for the reported issue.Further.Also a battery contact was loose from the circuit board.Electrical inspection could not be performed because of the observed malfunction.No injury was reported.This is a final report.
|
| |
|
Search Alerts/Recalls
|