EXACTECH, INC. EQUINOXE REVERSE SHOULDER COMPONENTS; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Model Number |
EQUINOXE REVERSE 42MM GLENOSPHERE |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
02/15/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant products: eq rev locking screw (cat# 320-15-05 / serial# (b)(4)); equinoxe, humeral stem primary, press fit 13mm (cat# 300-01-13 / serial# (b)(4)); equinoxe reverse tray adapter plate tray +0 (cat# 320-10-00 / serial# (b)(4)); eq reverse torque defining screw kit (cat# 320-20-00 / serial# (b)(4)); equinoxe reverse 42mm humeral liner +0 (cat# 320-42-00 / serial# (b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, approximately one months post initial right tsa, the 59 y/o male patient had a revision due to dislocation.Patient was too active after surgery.Patient was revised to exactech devices.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Sales rep was unable to obtain x-rays/photos.The devices are not available for evaluation as they were disposed by hospital.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3) as reported, approximately one months post initial right tsa, the 59 y/o male patient had a revision due to dislocation.Patient was too active after surgery.Patient was revised to exactech devices.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Sales rep was unable to obtain x-rays/photos.The devices are not available for evaluation as they were disposed by hospital.Additional information received indicates that there is no implant x-rays and any pre-revision x-rays for this revision or any additional information other than what was provided in the report.Based on review of all available information, there is no evidence to suggest that the reported dislocation and revision are related to any design, manufacturing, or patient related issues.The cause of the revision is most likely is related to the patient¿s underlying condition; however, that could not be confirmed.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, d4 (remove all from initial report-suspect products unknown d1, d4, d8, g4 (remove from initial report-501k unknown) h6 medical device problem code, type, findings and investigation conclusion.Based on review of all available information, there is no evidence to suggest that the reported dislocation and revision are related to any design, manufacturing, or patient related issues.The cause of the revision is most likely is related to the patient¿s underlying condition; however, that could not be confirmed.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|