| Model Number |
N/A |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: 110003450 g7 str monoblock shell insrtr, lot number 422210.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2023 - 00771 the product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the g7 cup was stuck to the insertion handle.The procedure was completed using another instrument.There is no additional information available at the time of this report.
|
| |
|
Event or Problem Description
|
|
There is no update to the prior event description provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the shell and inserter are assembled and could not be disassembled.The inserter shows scuffing on the shaft and taper near the handle.The strike plate of the inserter also shows impact marks.The shell has damage to the lip.Complaint confirmed based on returned product.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|