| Model Number |
67NFPS30 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Event problem and evaluation codes: updated.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product sample nor photos were received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon inflates lopsided when saline placed via cuff.Additional information received via customer response email 26-aug-2022: patient was not harmed, as trach was never placed.This trach's balloon is deformed, lopsided.
|
| |
|