| Model Number |
82087-01 |
| Medical Device Problem Codes |
Premature Activation (1484); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported activation/deployment failure was unable to be confirmed.The stent implant was exposed 1.3 millimeters from the distal end of the sheath.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As the reported difficulties were unable to be confirmed during return analysis, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that the distal shaft was restricted in the anatomy and/or with other devices preventing the shaft lumens from moving freely resulting in the reported activation/deployment failure.Manipulation of the device and/or handling during packing for return analysis likely resulted in the noted exposed stent implant.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported the procedure was to treat the left internal carotid artery with no noted calcification or tortuosity.The xact self expanding stent system (sess) was advanced to the lesion over a.014 non-abbott guide wire and a 7fr non-abbott sheath was used.The stent did not deploy at all when the thumbwheel was turned.There was no problem with the deployment mechanism.The sess was removed and a 10x8x30 and 7x7x30 xact stents were implanted without issue to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.Returned device analysis identified that the stent implant was exposed 1.3 millimeters (mm) from the distal end of the sheath.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|