| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed self test for defib and pacer functions.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device activity log.However, the device was put through extensive testing including functionality testing without duplicating the report.Internal inspection of the device found no discrepancies.No accessories were returned as part of this investigation.The processor/bridge/pace board was replaced and scrapped as a precaution.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|