• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT MEDICAL SHILEY; TUBE, BRONCHIAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT MEDICAL SHILEY; TUBE, BRONCHIAL (W/WO CONNECTOR) Back to Search Results
Model Number 125037
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, both device's cuff had an inflation or deflation issue.There was no patient involvement.
 
Additional Manufacturer Narrative
Concomitant medical product: 125037 37fr lt bronco cath pu cuff x1, (lot#: unknown).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that there is a large slit in the main body of the tracheal cuff and there is stick tape attached to the main stem of the tubing.Functionally, an attempt made to inflate the returned unit shows that the bronchial cuff inflated without any issue however the tracheal cuff failed to inflate.Further examination of the tracheal cuff body shows that there is a ¿v¿ shaped slit in the main body of the cuff.It was reported that the device's cuff had an inflation or deflation issue.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the cuff body came in contact with a sharp utensil or object.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: various bony anatomical structures (e.G., teeth) within the intubation route or any intubation tools with sharp surfaces present a threat to maintaining cuff integrity.Care must be taken to avoid damaging the thin-walled cuffs during intubation which would create the need to subject the patient to the trauma of extubation and re-intubation.If either cuff is damaged, the tube should not be used.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, BRONCHIAL (W/WO CONNECTOR)
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
EI  3810
Manufacturer (Section G)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
EI   3810
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16702745
Report Number8020889-2023-00067
Device Sequence Number5669659
Product Code BTS
UDI-Device Identifier10884521831568
UDI-Public10884521831568
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K771219
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number125037
Device Catalogue Number125037
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2023
Supplement Date Received by Manufacturer04/26/2023
Initial Report FDA Received Date04/07/2023
Supplement Report FDA Received Date05/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-