• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. GRAY RELOAD FOR 45MM ECHELON; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. GRAY RELOAD FOR 45MM ECHELON; STAPLE, IMPLANTABLE Back to Search Results
Model Number ECR45M
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4) date sent: 4/7/2023 d4: batch # 400a88 investigation summary the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample determined that one ecr45m reload was received partially fired 1/10 and with an opened package.The returned reload was reset and loaded into a test device.The device was tested for functionality in the straight position and achieved its complete stroke firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples meet the staple form release criteria.It is possible that while loading the reload, the reload was pushed farther back than the reload alignment slot resulting in the knife pushing the one-piece sled forward and locking the reload.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 400a88, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during the surgery, after fired, noted the staples malformed.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAY RELOAD FOR 45MM ECHELON
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key16703364
Report Number3005075853-2023-02412
Device Sequence Number3640847
Product Code GDW
UDI-Device Identifier10705036001751
UDI-Public10705036001751
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/30/2026
Device Model NumberECR45M
Device Catalogue NumberECR45M
Device Lot Number400A88
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/20/2023
Initial Report FDA Received Date04/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-