| Model Number |
TSM-240 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Health effect - impact code: f2201 biopsy.A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown and rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture, "recalled", and capsular contracture baker grade unknown.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: capsular contracture: unable to observe as it is a medical event not related to the device.Rupture: no observed rupture on the device.Additional observations: crease fold observed on the device.White particles observed on the device.Wear abrasion and crease fold observed on the device.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture baker grade unknown.Later healthcare professional reported right side rupture.Device has been explanted.
|
| |
|