| Model Number |
25MTJ-503 |
| Medical Device Problem Code |
Perivalvular Leak (1457)
|
| Health Effect - Clinical Codes |
Endocarditis (1834); Fever (1858)
|
| Date of Event |
03/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on 13 january 2018, a 25mm sjm masters series heart valve with teflon cuff was successful implanted.On (b)(6) 2023, the patient presented with infective endocarditis and a re-do procedure was performed.There was no particular problem due to it being a mechanical valve, but reoperation was performed because of para-valvular leak.The device was explanted and replaced with another 25mm sjm masters series heart valve with teflon cuff.No additional information was reported.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, a 25mm sjm masters series heart valve with teflon cuff was successful implanted.On (b)(6) 2023, the patient presented with infective endocarditis and a re-do procedure was performed.There was no particular problem due to it being a mechanical valve, but reoperation was performed because of para-valvular leak.The device was explanted and replaced with another 25mm sjm masters series heart valve with teflon cuff.Subsequent to the previously filed report, additional information was received that the patient had a fever due to the infective endocarditis.The patient does not have any history of endocarditis.The patient does not have a history an indwelling vascular catheter, intravenous drug use, recent dental work, or injury from a non-sterile object.The paravalvular leak is believed to have been a result of the infective endocarditis.The implanting physician believes that the cause of the endocarditis is due to a patient condition.The patient was stable at the time of report.
|
| |
|
Additional Manufacturer Narrative
|
|
Explant due to perivalvular leak and endocarditis was reported.The investigation found that both leaflets were intact and opened and closed completely without difficulty.However, part of the sewing cuff was removed and what remained contained a small amount of tissue.There was no allegation of malfunction against the abbott device.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received the cause of the reported incident could not conclusively be determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|