| Model Number |
305917 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Examination of the product involved may provide clarification as to the cause for the reported failure.We appreciate you taking the time to bring this observation to our attention.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd safetyglide¿ needle only, 21 g there was a crack.There was no report of patient impact.The following information was provided by the initial reporter: it was reported that the bd safetyglide¿ needle "cracked, broken tip or tip strength" the following information was provided by the initial reporter: cracked, broken tip or tip strength <5kg (11lb), applied to 3mm from the top of.The tip ¿ aql 0.15.Broken or unusable luer lock adaptor (lla) ¿ aql 0.015.Missing lla ¿ aql 0.065.
|
| |
|