MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-712 PARADIGM INSULIN PUMP; PUMP, INFUSION, INSULIN
|
Back to Search Results |
|
Model Number MMT-712WWS |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/23/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Select patient information cannot be provided due to regional privacy regulations.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Information received by medtronic indicated that the customer reported pump led screen damage and it was blurred and hard to read.No harm requiring medical intervention was reported.Troubleshooting was not performed.It was unknown that the customer will continue using the pump but pump will not be returned for analysis.
|
|
Manufacturer Narrative
|
Unit received with missing segment/partial display due to cracked and bleeding lcd glass.Unable to perform displacement test due to missing segment/partial display.No moisture damage found on the electronics, motor, battery tube and vibrator assembly noted.The test reservoir locked in place properly and no anomaly noted.The following were noted during visual inspection: cracked reservoir tube lip, cracked battery tube threads, stained keypad overlay, stained address/serial number label and stained end cap sticker.Unit received missing segment/partial display due to cracked and bleeding lcd glass confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Updated summary: the insulin pump was returned for analysis.
|
|