• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 CORAIL2 LAT COXA VARA SIZE 10; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 CORAIL2 LAT COXA VARA SIZE 10; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 3L93710
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 01/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).The initial reporter has been removed for confidentiality/privacy.The initial reporter is the patient.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient initiated complaint as follows: i am reaching out to share what if feel is really important information with you.My journey as a hip replacement surgery began in (b)(6) 2018 when i had my initial right hip surgery.I recovered very well no issues until (b)(6), 2023 (b)(6), 2023, my hip began to "squeak" like an old screen door.I was unable to walk without experiencing pain and every step induced a squeaky step.My original surgeon has since retired so the next available appointment was wednesday (b)(6) at which time it was decided that there was a malfunction to my hip replacement which is less than 5 years old.(b)(6) 2023, i have a re-do of my original hip replacement spent two days inpatient and followed up with post op physical therapy.I did follow up with my routine 2 week appointment and all was going well.Doi: (b)(6), 2018.Dor: (b)(6), 2023.Affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.The photographs and video attached were reviewed.However, they do not represent the current complaint condition and event.Therefore, the investigation could not draw any conclusions about the reported event.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed, for the finished device lot#: 5301961.And no nonconformances were identified.
 
Event or Problem Description
Additional information received: on (b)(6) 2018 patient underwent a right tha.On (b)(6) 2023: patient began experiencing squeaking (no pain).On (b)(6) 2023: the patient underwent a revision for squeaking.(b)(4) on (b)(6) 2023: patient dislocated while getting up from a couch and experienced pain.The patient underwent a closed reduction under sedation on (b)(6) 2023.The patient was placed in a leg brace due to continued instability which indicated the need for another revision.On (b)(6) 2023: the patient underwent a second revision.(b)(4).
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Removed date of explant since there was no indication that the stem was revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 LAT COXA VARA SIZE 10
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16706868
Report Number1818910-2023-07552
Device Sequence Number11672730
Product Code LZO
UDI-Device Identifier10603295168874
UDI-Public10603295168874
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K192946
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number3L93710
Device Catalogue Number3L93710
Device Lot Number5301961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2023
Supplement Date Received by Manufacturer05/03/2023
05/25/2023
Initial Report FDA Received Date04/10/2023
Supplement Report FDA Received Date05/09/2023
05/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/01/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX NEUT 32IDX50OD; DELTA CER HEAD 12/14 32MM +1; PINN CAN BONE SCREW 6.5MMX25MM; PINNACLE SECTOR II CUP 50MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexFemale
-
-