Model Number 86-7428
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 03/16/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Revision of a fractured cemented distal femoral stem left side.Doi: unknown dor: (b)(6) 2023.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional information received indicates that the patient was experiencing instability that led to the revision surgery and the initial implantation took place in another hospital in 2014.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D10 (updated the concomitant product unknown bone cement to palacos cement (competitor).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned to depuy synthes for evaluation.Visual examination found the device broken at the shaft near the distal base of the stem, confirming the reported allegation.Additionally, heavy extraction damage can be identified.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: 1) quantity manufactured: (b)(4).2) date of manufacture: 4/17/14.3) any anomalies or deviations identified in dhr: none.4) expiry date: 2024/02.5) ifu reference: 092-00-252 revk.Device history review: 1) quantity manufactured: (b)(4).2) date of manufacture: 4/17/14.3) any anomalies or deviations identified in dhr: none.4) expiry date: 2024/02.5) ifu reference: 092-00-252 revk.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LPS CEM FEM STEM 11X200MM BOW
Common Device Name
UNIVERSAL KNEE STEMS AND SLEEVES : KNEE STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16708866
Report Number1818910-2023-07583
Device Sequence Number18005892
Product Code JWH
UDI-Device Identifier10603295078807
UDI-Public10603295078807
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K003182
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number86-7428
Device Catalogue Number198715311
Device Lot Number479256
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2023
Supplement Date Received by Manufacturer05/04/2023
06/13/2023
07/05/2023
Initial Report FDA Received Date04/10/2023
Supplement Report FDA Received Date05/10/2023
06/27/2023
07/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ETHISORB MEDULLARY PLUG UND; LPS CEM FEM STEM 11X200MM BOW; LPS DISTAL FEM COMP XSM LT; LPS UNIV TIB HIN INS XSM 12MM; MBT REV TIB TRAY SIZE 4 15MM; MBT TRAY SLEEVE POR M/L 53MM; PALACOS CEMENT (COMPETITOR); UNIVERSAL STEM 115X14MM FLUTED; UNKNOWN BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-