| Model Number |
86-7428 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Joint Laxity (4526)
|
| Date of Event |
03/16/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision of a fractured cemented distal femoral stem left side.Doi: unknown dor: (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received indicates that the patient was experiencing instability that led to the revision surgery and the initial implantation took place in another hospital in 2014.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D10 (updated the concomitant product unknown bone cement to palacos cement (competitor).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned to depuy synthes for evaluation.Visual examination found the device broken at the shaft near the distal base of the stem, confirming the reported allegation.Additionally, heavy extraction damage can be identified.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: 1) quantity manufactured: (b)(4).2) date of manufacture: 4/17/14.3) any anomalies or deviations identified in dhr: none.4) expiry date: 2024/02.5) ifu reference: 092-00-252 revk.Device history review: 1) quantity manufactured: (b)(4).2) date of manufacture: 4/17/14.3) any anomalies or deviations identified in dhr: none.4) expiry date: 2024/02.5) ifu reference: 092-00-252 revk.
|
| |
|
Search Alerts/Recalls
|