| Catalog Number |
06481647190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lamp was replaced on 01-nov-2022; the cuvettes were replaced "more recently." it was noted that service "has not been performed for a long time." the customer declined to provide any additional information.Based on the information provided, the investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|
Event or Problem Description
|
|
The initial reporter complained of discrepant results for 1 patient sample tested for mg2 magnesium gen.2 (mg2) on a cobas 4000 c (311) stand alone system.The initial result from the c311 system was 0.9 mg/dl.The repeat result from the same system was 1 mg/dl.The sample was sent to an external laboratory and the result from an unknown method was 1.9 mg/dl.No questionable results were reported outside of the laboratory.The c311 analyzer serial number was not provided.
|
| |
|