• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS MG2; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS MG2; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Catalog Number 06481647190
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lamp was replaced on 01-nov-2022; the cuvettes were replaced "more recently." it was noted that service "has not been performed for a long time." the customer declined to provide any additional information.Based on the information provided, the investigation did not identify a product problem.The cause of the event could not be determined.
 
Event or Problem Description
The initial reporter complained of discrepant results for 1 patient sample tested for mg2 magnesium gen.2 (mg2) on a cobas 4000 c (311) stand alone system.The initial result from the c311 system was 0.9 mg/dl.The repeat result from the same system was 1 mg/dl.The sample was sent to an external laboratory and the result from an unknown method was 1.9 mg/dl.No questionable results were reported outside of the laboratory.The c311 analyzer serial number was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MG2
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16709615
Report Number1823260-2023-01199
Device Sequence Number12626111
Product Code JGJ
UDI-Device Identifier04015630929818
UDI-Public04015630929818
Combination Product (Y/N)N
Initial Reporter CountryGR
PMA/510(K) Number
K954992
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Catalogue Number06481647190
Device Lot Number63779901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2023
Initial Report FDA Received Date04/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexMale
-
-