SMITH & NEPHEW ORTHOPAEDICS AG TRIAL FEMORAL HEAD 36 S/+0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
|
Back to Search Results |
|
| Model Number |
75100856 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a thr, the inner locking mechanism that attaches to a trial femoral head 36 s/+0 had fractured.Furthermore, no pieces fell into the patient.Surgery was resumed, without any delay, with a smith and nephew back-up device.Patient was not harmed as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that, during a total hip replacement, the inner locking mechanism that attaches to a trial femoral head 36 s/+0 had fractured.The device intended for use in treatment was not returned for investigation.A product evaluation was not possible.No batch number was communicated so the document history review was not possible.Due to an unknown batch number, the review of historical complaints was performed on product number basis only, revealing 10 additional complaints over the past 12 months with similar failure mode.Due to insufficient information it is not possible to perform a review of past corrective actions.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.The performed investigation does not lead to an accurately determined cause.Based on the available information it is not possible to investigate whether the reported device met manufacturing specifications upon release for distribution.Due to insufficient information, it is not possible to speculate about factors which could have contributed to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.There is no need for further actions because of the limited information provided.Nevertheless, smith + nephew will continue to monitor this device for similar issues.This complaint will be reopened should additional information or the device be received.
|
| |
|
Search Alerts/Recalls
|
|
|