Model Number FR8A-TRL-A0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The surgical issue questionnaire was reviewed for potential causes of the reported issue.The clinical representative did not notice the expiration date and (b)(4) was opened to investigate the expired product being implanted.A representative conducted a review of sterilization and packaging records for the respective product lot; it has been confirmed that the product was delivered sterile, validated sterilization parameters were used, and sterile barriers were verified to be intact following packaging.The stimulator is used to treat pain.The cause of implanting an expired product is due to the clinical representative providing the clinician with an expired trial stimulator (user error - clinical representative).Refer to (b)(4) for investigation details, including risk assessment of implanting an expired stimulator.
 
Event or Problem Description
An expired freedom scs neurostimulator trial lead was implanted.The product was shipped to the clinical representative on june 18, 2021, expired on march 01, 2023 and was implanted on (b)(6) 2023.The lead pull was performed as expected on (b)(6) 2023.The patient had a successful trial with no adverse events reported, and will be moving forward with a permanent implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEDOM SPINAL CORD STIMULATOR
Common Device Name
SPINAL CORD STIMULATOR
Manufacturer (Section D)
STIMWAVE TECHNOLOGIES INC.
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer (Section G)
STIMWAVE TECHNOLOGIES INC.
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer Contact
alicia pagliaro
1310 park central boulevard s.
pompano beach, FL 33064
8009655134
MDR Report Key16709729
Report Number3010676138-2023-00044
Device Sequence Number5920883
Product Code GZB
UDI-Device Identifier00818225020679
UDI-Public(01)00818225020679(17)230301
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K182720
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date03/01/2023
Device Model NumberFR8A-TRL-A0
Device Lot NumberSWO210301
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2023
Initial Report FDA Received Date04/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-