| Model Number |
IPN926926 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "the cuff wouldn't inflate".The customer was unable to answer our multiple inquiries regarding if the event happened while in use on a patient, how long was the device in use, if medical intervention was required, if reintubation was required, if desaturation occurred, and what the current health status of the patient is.If additional information is received, the complaint file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that "the cuff wouldn't inflate".The customer was unable to answer our multiple inquiries regarding if the event happened while in use on a patient, how long was the device in use, if medical intervention was required, if reintubation was required, if desaturation occurred, and what the current health status of the patient is.If additional information is received, the complaint file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer returned one unit 5-10115 et tube, cf, 7.5 for investigation.One representative sample was returned.An actual sample was not returned.The representative sample was visually examined with and without magnification.Visual examination of the returned device revealed that the sample appears typical.No defects or anomalies were observed.A functional inspection was performed to attempt to inflate and deflate the balloon cuff on the returned et tube.A lab inventory syringe filled with air was connected to the valve of the pilot balloon.Upon injection of air, both the pilot and cuff immediately inflated.The syringe was then removed and the cuff and pilot balloon were inspected for leaks.No leaks were detected.The syringe was then reattached in order to deflate the balloons.Both the pilot balloon and balloon cuff were able to deflate.The sample was submerged underwater to test for any leaks.However, no leaks were detected.No issues were found with the returned sample as it was able to properly inflate and deflate.A device history record review was performed and no relevant findings were identified.The ifu for this product states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is a detected in any part of the system , the tube should not be used." "care should be taken to avoid damaging the thin-walled cuffs during intubation.If cuff is damaged, the tube should not be used." the ifu also states, "cuff pressure should be monitored routinely to assure that an adequate tracheal seal is maintained and that the cuff has not become over-inflated.The tracheal tube cuff should be slowly inflated with the minimum amount of air required to provide an effective tracheal seal.Do not inflate with a measured volume of air or by "feel" of pressure in the syringe since little resistance should be felt during inflation." the reported complaint could not be confirmed since the actual sample was not returned.The customer returned a representative sample in place of the actual sample that resulted in the complaint issue.Upon functional inspection, no problems were found with the returned device as it was able to properly inflate and deflate.No leaks were found with the sample.A device history record review was performed on the device with no evidence to suggest a manufacturing related cause.The probable cause of this complaint issue could not be determined without the actual device.Other remarks: n/a.Corrected data: n/a.
|
| |
|