It was reported that patient had surgery on september 8, 2022 and was overcorrected on both eyes.Right eye (od) - idesign manifest refraction (mr) -2.96 +0.52 x 29, mr -2.91 +.25 x 10 20/20 best corrected visual acuity (bcva).Left eye (os) - idesign mr -2.93 +0.18 x 172, mr -3.16 sph.20/20 bcva.Enhancement done on december 22, 2022 for left eye only.Idesign mr +2.54 +1.45 x 0, mr +1.84 +1.00 x 10.Physician adj -0.50.As of (b)(6) 2023 the patient was seeing 20/30 uncorrected visual acuity (ucva), they have not done an mr on the patient.While the od was not enhanced, the current mr is +1.25 +.50 x15 with 20/20 bcva.However, the patient wears a contact lens (cl) and only sees 20/30 with the cl.Patient declined an enhancement on the od.The patient does appear to have a good bit of higher order aberrations (hoas) in both eyes.The idesign se for the os was -2.84 and the eff blur was 4.14 d.Pre-op k¿s for the os were 45.30/46.57 x168 and after the enhancement were measuring 39.30/41.90 x175.No additional information was provided.This report is for the idesign.A separate report will be filed for the excimer laser.
|