| Model Number |
15 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device did not recognize the therapy cable.In this state, the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported during the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer¿s device and was unable to verify or duplicate the reported issue.The device passed functional and performance testing and was returned to the customer for use.The cause of the reported issue could not be determined.
|
| |
|