ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER EDGEPLUS VALVED ENTRY SYSTEM; CANNULA, TROCAR, OPHTHALMIC Back to Search Results
Catalog Number 8065751657
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A government agency reported that during vitrectomy surgery water kept coming out from an ophthalmic trocar.The surgery was completed after replacing the product with another one.There was no patient harm.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation for the report of during surgery water kept coming out from the trocar and was replaced another one; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The cause of the reported event cannot be determined with the information obtained, therefore, specific action with regards to this complaint cannot be taken.An acceptance quality inspection is performed to ensure product meets release acceptance criteria.This inspection includes evaluation for damaged valves.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGEPLUS VALVED ENTRY SYSTEM
Common Device Name
CANNULA, TROCAR, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16711387
Report Number2028159-2023-00481
Device Sequence Number7274535
Product Code NGY
UDI-Device Identifier00380657516575
UDI-Public00380657516575
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2024
Device Catalogue Number8065751657
Device Lot Number14CH9C
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/13/2023
Supplement Date Received by Manufacturer05/03/2023
06/09/2023
Initial Report FDA Received Date04/11/2023
Supplement Report FDA Received Date05/29/2023
07/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/02/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-