| Model Number |
IPN923940 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#: (b)(4).Other remarks: n/a; corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that while in use on a patient, when "the [operator] went to remove the inner guidewire, the blue end fell off.They were able to get the wire out, but the catheter unraveled and kinked towards the tip.They replaced the catheter with another and the same thing happened".No patient harm or injury.The patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that while in use on a patient, when "the [operator] went to remove the inner guidewire, the blue end fell off.They were able to get the wire out, but the catheter unraveled and kinked towards the tip.They replaced the catheter with another and the same thing happened".No patient harm or injury.The patient status is reported as "fine".
|
| |
|
Search Alerts/Recalls
|