Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
An endurant iis bifurcate stent graft system was implanted in the endovascular treatment of a 53mm aaa. it was reported that the patient returned for follow-up , a type ia endoleak was diagnosed. on the same date , intervention was performed.The physician first implanted a non mdt cuff followed by an endurant cuff and endoanchors.The patient had prior non mdt stents placed in the limbs of the existing graft.During the advancement of one of the devices, one of the non mdt stents was unknowingly pushed proximally and was trapped by the suprarenal fixation on the endurant cuff after it was deployed.This did not affect the deployment of the endurant cuff or cause it to be implanted outside of the intended area.Per the physician the cause of the type ia endoleak can not be determined. the cause of the issue with the endurant cuff trapping a stent by its suprarenal fixation is user error.No additional clinical sequalae were provided and the patient is fine.
|