| Model Number |
H7493941832300 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.The synergy xd mr us 3.00 x 32mm stent delivery system was returned for analysis.The stent profile and tip profile section of the delivery system were not returned.A visual and tactile examination of the hypotube shaft found multiple kinks along several locations of the hypotube shaft.A shaft break was also noted at 120 cm distal to the distal end of the strain relief.A visual and tactile examination of the shaft polymer extrusion profile could not be performed as this section was not returned.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2023.It was reported that stent unable to load the guidewire.A 3.00 x 32mm synergy xd stent was selected for use.However, during insertion the device failed to load the 0.014 guidewire.The device was removed and the procedure was completed successfully with a new 3.0x32 synergy xd stent.There were no patient complications nor injuries reported.However, returned device analysis revealed a hypotube detachment and stent was not returned.
|
| |
|