This follow-up is being submitted to relay additional information.Updated: b4, b5, d2, d9, g1, g3, g6, h1, h2, h3, h6, and h10 visual evaluation of the returned product found the inner sterile pouches are broken into multiple pieces.Evaluation of the outer pouch found no damage.Sterility has not been breached.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event cannot be determined.If any further information is received which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|