• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ 10 5MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ 10 5MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-21-005
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that locking mechanism did not work.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: the device associated with this report was not received for examination.Per the photographic evidence it was not possible to visualize any defect and/or anomaly which could affect to the proper assembly of the device.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: quantity manufactured: (b)(4).Date of manufacture: 11/19/2021.Any anomalies or deviations identified in dhr: none.Expiry date: 10/31/2026.Ifu reference: 090200828.Device history review: a manufacturing record evaluation was performed for the finished device (151621005/jn8193) product and lot numbers, and no non-conformances were identified.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9 and d10 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: the device associated with this report was returned to depuy synthes for evaluation.Visual examination of the returned tibial insert found evidence of deformation/witness marks along one of the posterior grooves that is designed to slide into the locking feature of the mating tibial tray.The observed damage is consistent with unsuccessful insertion attempts during the procedure.The mating tibial tray was not returned, therefore, a functional test was not able to be performed.However, due to the visual damage observed, the difficulty/inability to assemble mating devices can be confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: 1) quantity manufactured: (b)(4) pcs; 2) date of manufacture: 11/19/2021; 3) any anomalies or deviations identified in dhr: none; 4) expiry date: 10/31/2026; 5) ifu reference: (b)(4).Device history review: 1) quantity manufactured: (b)(4) pcs; 2) date of manufacture: 11/19/2021; 3) any anomalies or deviations identified in dhr: none; 4) expiry date: 10/31/2026; 5) ifu reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FB INSRT SZ 10 5MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16715856
Report Number1818910-2023-07720
Device Sequence Number7263797
Product Code JWH
UDI-Device Identifier10603295046240
UDI-Public10603295046240
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1516-21-005
Device Catalogue Number151621005
Device Lot NumberJN8193
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2023
Supplement Date Received by Manufacturer05/22/2023
05/31/2023
06/21/2023
Initial Report FDA Received Date04/11/2023
Supplement Report FDA Received Date05/23/2023
06/01/2023
06/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN KNEE TIBIAL TRAY.
Patient SexUnknown
-
-