STRYKER ENDOSCOPY-SAN JOSE PNEUMOCLEAR PLUS CO2 CONDITIONING INSUFFLATOR (AMERICAS); INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Model Number FM300
Medical Device Problem Code Pressure Problem (3012)
Health Effect - Clinical Code Unspecified Respiratory Problem (4464)
Date of Event 03/15/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient suffered from subcutaneous emphysema resulting in the patient needing a chest tube and icu care.
 
Additional Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.Filling on behalf of world of medicine (wom).
 
Additional Manufacturer Narrative
The product was not returned for investigation therefore the reported failure mode was not confirmed.Alleged failure: pressure issue.Expanded investigation activities.The device was investigated by stryker on-site with the following conclusion provided on march 31,2023.The device is overdue for its scheduled 2-year preventative maintenance; however, no error logs were present, and the pressure sensor is displaying correct values.The "exceed safety threshold" alert functioned correctly in all operating modes.The "exceed safety threshold" alert was displayed during the case with the alleged incident.Probable root cause: because device was not returned to oem - wom, probable root cause cannot be determined.The device was investigated by the distributor at the end users facility.The results of the functional inspection were as follows: the device is overdue for its scheduled 2-year preventative maintenance; however, no error logs were present, and the pressure sensor is displaying correct values.The "exceed safety threshold" message and signal functioned correctly in all operating modes.The "exceed safety threshold" message was displayed during the case with the alleged incident.No further action is indicated at this time.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Event or Problem Description
It was reported that the patient suffered from subcutaneous emphysema resulting in the patient needing a chest tube and icu care.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUMOCLEAR PLUS CO2 CONDITIONING INSUFFLATOR (AMERICAS)
Common Device Name
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16718655
Report Number0002936485-2023-00293
Device Sequence Number11174962
Product Code HIF
UDI-Device Identifier04056702001308
UDI-Public04056702001308
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K170784
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFM300
Device Catalogue Number0620050000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/16/2023
Supplement Date Received by Manufacturer03/16/2023
Initial Report FDA Received Date04/11/2023
Supplement Report FDA Received Date01/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/15/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;