| Model Number |
CAT01858 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke during procedure.All pieces were retrieved.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke during procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: nanotack 1.4 flex anchors not following path of drill, in turn there was bending of the anchors.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be: 1) guide out of specification, 2) pathology such as schlerotic bone that may thicken the cortical layer.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|