Model Number 02.18.IF3.09.LM
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 03/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 21 march 2023: lot 2207120: (b)(4) manufactured and released on 15-jun-2022.Expiration date: 2027-may-19.No anomalies found related to the problem.To date, (b)(4) of the same lot have been sold with no similar reported event during the period of review.
 
Event or Problem Description
The patient came in due to signs of infection and the pathogen was unknown.About 2 months after the primary surgery, the surgeon performed a washout and poly swap and the surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTO PARTIAL KNEE TIBIAL INSERT FIX S3 LM - 9MM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key16723120
Report Number3005180920-2023-00242
Device Sequence Number12560389
Product Code HSX
UDI-Device Identifier07630030895999
UDI-Public07630030895999
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K162084
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.18.IF3.09.LM
Device Catalogue Number02.18.IF3.09.LM
Device Lot Number2207120
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2023
Initial Report FDA Received Date04/12/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexMale
Patient Weight88 KG
-
-