|
|
| Model Number |
02.231.675 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Pain (1994)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional pro-code: hrs, jdp.Complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter is lawyer.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, the patient underwent left total knee arthroplasty with distal replacement.Hardware removal left femur due to left supracondylar/intercondylar femoral fracture nonunion, hardware failure and osteoarthritis left knee.On (b)(6) 2021 the patient underwent left distal femur open reduction and internal fixation with a synthes lateral plate and multiple cerclage cables due to left distal femur comminuted intra-articular fracture and morbid obesity.It was reported that on (b)(6) 2021, the patient had a fall.She was climbing over something like a dog gate, tripped and fell.She initially felt pain in her left knee.She was diagnosed with a left distal femur fracture.This complaint involves seven (7) devices.This report is for one (1) 5.0mm cannulated va locking screw/75mm.This is report 5 of 7 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Photo investigation: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device images.Visual analysis of the photo revealed that the 5.0mm cannulated va locking screw/75mm had no defect found on it.Photos provided does not show any evidence to confirm any defect on the device.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed for the 5.0mm cannulated va locking screw/75mm.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.There is no device problem associated with this complaint.This complaint is adverse event only.
|
| |
|
Search Alerts/Recalls
|
|
|