• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number SSLP-360
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been/will be initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of infection is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.The reason for reoperation is: "developed mrsa".
 
Event or Problem Description
Healthcare professional reported "developed mrsa in the right breast implant exchange/capsulectomy side requiring the removal of her implant so this surgery was to replace it." the device has been explanted and replaced.
 
Event or Problem Description
Healthcare professional reported "developed mrsa in the right breast implant exchange/capsulectomy side requiring the removal of her implant so this surgery was to replace it." the device has been explanted and replaced.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: h10 fi summary: with the information collected during the investigation, there is enough evidence to support that device serial number was manufactured under controlled conditions in accordance with abbvie¿s procedures and were no defects/problems/issues found in the documents related to the reported event.Seal strength results were acceptable.Environmental monitoring results were found to be acceptable, and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.The sterilization run is not related to any additional complaint infection record reported as of today.During the trend review of all gel infection complaints no adverse trend was noted.Given the fact that the cause of the infection cannot be specifically associated to the manufacturing process, and there is not an adverse trend for this type of event, and no er/ ncr(s) or temporary changes were identified during the investigation associated to this lot and the complaint, no corrective action is deemed necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key16724364
Report Number9617229-2023-06489
Device Sequence Number20639567
Product Code FTR
UDI-Device Identifier10888628032767
UDI-Public10888628032767
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSSLP-360
Device Catalogue NumberSSLP-360
Device Lot Number3596141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2023
Supplement Date Received by Manufacturer04/20/2023
Initial Report FDA Received Date04/12/2023
Supplement Report FDA Received Date05/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-