• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS CKMBL CREATINE KINASE-MB; CPK OR ISOENZYMES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS CKMBL CREATINE KINASE-MB; CPK OR ISOENZYMES Back to Search Results
Catalog Number 07190794190
Medical Device Problem Codes Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer simulated the situation again and a data flag was present on the sample.Based on the provided data and information, the investigation determined the event was consistent with an issue related to the sample concentration.A possible analyzer contribution regarding one sample without an alarm could not be investigated due to missing reaction kinetics.Based on the quality control data, a general problem with the reagent could be ruled out.
 
Event or Problem Description
There was an allegation of a questionable creatinine kinase (ck) result from the cobas 6000 c501 module serial number (b)(4).In the afternoon at 18:04, the initial result for sample 1 was 733 u/l (with no data flag) and was reported outside of the laboratory.Later in the evening around 22:00, sample 2 was drawn from the patient and the ck result was 2505 u/l with a data flag.With a 1:50 dilution, the result was 34941 u/l.Sample 1 was then repeated at 22:45 with a 1:50 dilution and the result was 37337 u/l.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CKMBL CREATINE KINASE-MB
Common Device Name
CPK OR ISOENZYMES
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16726441
Report Number1823260-2023-01230
Device Sequence Number7209581
Product Code JHS
UDI-Device Identifier04015630940806
UDI-Public04015630940806
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K951595
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Catalogue Number07190794190
Device Lot Number66484301
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2023
Initial Report FDA Received Date04/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-