| Model Number |
133S-FX-13-T |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Dhr trend summary: no patterns were found; therefore, no additional actions are deemed required.The trend of a050401 - fluid/blood leak complaints will continue to be monitored and corrective actions will be taken in the future if deemed appropriate.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: device deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side deflation.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d.9., h.3., h.6.Device evaluation based on the device analysis grid, the assessments of the complaint are: ¿ deflation : observed two opening on the radius side assessed as surgical damage and more than three opening on the anterior side assessed as surgical damage like.No additional observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side deflation.Device has been explanted.
|
| |
|