|
|
| Model Number |
1663G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).H6.Component code: g07002.Device not returned.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The single complaint was reported with multiple events.There are no additional details regarding the additional events.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Could you please clarify if the patient suffered from any signs or consequences due to the issue breakage suture? please provide more information.A manufacturing record evaluation was performed for the finished device lot, and no non conformances were identified.Event related to mw # 2210968-2023-02653, mw # 2210968-2023-02656, mw # 2210968-2023-02659, mw # 2210968-2023-02660, mw # 2210968-2023-02661.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.The suture kept breaking on surgeon who said it was falling apart.As they were tying the suture it was breaking.They mentioned that this was happening several times out of this box of suture.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional h6 component code: g07002 reported condition not confirmed.Additional h3 investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional test evaluation were conducted on the returned device.The returned sample revealed that twelve packets that pertain to product code 1663g were returned for analysis.Upon initial inspection, of the samples, no external damages were observed on the packets.In order to evaluate the condition of the returned samples, the packets were opened.The swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to damaged or breakage sutures and no defects were observed during evaluation.A functional test was performed, and the tensile strength results were above the minimum requirements.The device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|