| Model Number |
70-1028-0915 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cage would not expand and the surgeon ended up having to take out the implant.They tried opening a second cage and the same issue occurred.The implant only expanded with extreme downward pressure.The ended up explanting the second one as well and going with a straight aleutian cage.There was a 60 min surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Without the device we are unable to determine true cause, but upon discussion it appears the implant was packed with graft prior to expansion which is not the correct method.
|
| |
|