Model Number 750180
Medical Device Problem Codes Loss of or Failure to Bond (1068); Deflation Problem (1149); Material Too Soft/Flexible (4007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the cannula could not be blocked.It was blocked with 12 milliliters (ml) of air to begin with.There was a kink/adhesion on the side of the cuff balloon and the cuff could not be completely blocked properly.After repeated unblocking and blocking, the adhesion came off.However, at 28-30 bar, the cuff and test balloon were very soft and misshapen.In the end, unblocking could not be done completely with a cuff manager.Residual air in the cuff had to be unblocked with a syringe.With the cannula in place, this could have led to severe irritation or even damage to the patient.No injury was reported.
 
Additional Manufacturer Narrative
Other text: h3 and h6 - evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection found no unusual condition was detected in the sample.No root cause could be determined since the complaint was not confirmed; the sample provided does not show any failure mode.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.No corrective actions are required since the complaint was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE, ADULT
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16728337
Report Number3012307300-2023-04054
Device Sequence Number12437583
Product Code JOH
UDI-Device Identifier15021312006421
UDI-Public15021312006421
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K083641
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/17/2025
Device Model Number750180
Device Catalogue Number750180
Device Lot Number4033788
Was Device Available for Evaluation? No
Date Returned to Manufacturer04/03/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2023
Supplement Date Received by Manufacturer08/08/2023
Initial Report FDA Received Date04/12/2023
Supplement Report FDA Received Date08/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/08/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-