|
Model Number LTF-S190-10 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/06/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
E1: (b)(6).The device was returned to olympus for evaluation.The customer's complaint was confirmed.And likely, due to a cut on the charge coupled device cable.In addition to what's reported in b5, the following nonreportable malfunctions were identified: due to damage on light guide cover glass; water tightness was lost.Scratches on various parts of the device, bending section cover adhesive had a chip.Due to wear of angle wire; bending angle in up direction does not meet the specification.Connector leakage, switch button discoloration, and the angle wired elongated.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
The customer reported to olympus, that there was a "disconnection" when using the endoeye flex deflectable videoscope.The issue was identified during preparation for use for a therapeutic procedure.The device was inspected before use.The patient did not have any implants, and the procedure was completed.There was no patient harm associated with the event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.It is likely that the internal board for processing the video signal was damaged for some reason, resulting in a communication error with the video processor.The inspection method for the event is described as follows in the instructions for use (ifu) "chapter 3 preparation and inspection 3.8 checking the function in combination with related equipment"."[inspection of endoscopic images] check if normal light observation images and nbi observation images are displayed normally.1.Wipe the endoscope tip lens with sterile gauze moistened with saline or sterile water before inspection.2.Observe the palm, etc.With normal light observation images and nbi observation images.3.Make sure the light is coming out.(see figure 3.23) 4.Adjust the amount of light to get the proper brightness.5.Check that there are no abnormalities such as noise, blur, or cloudiness in the normal light observation image and the nbi observation image.6.Operate the angle lever of the endoscope to bend the curved part.7.Check that normal light observation images and nbi observation images do not disappear for a moment or other abnormalities." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|