| Catalog Number |
303367 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Needle Stick/Puncture (2462); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd interlink¿ vial access cannula broke at the base.The following information was provided by the initial reporter: the cannula broke at the base then punctured the web of her left hand going through and through.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd interlink¿ vial access cannula broke at the base.The following information was provided by the initial reporter: the cannula broke at the base then punctured the web of her left hand going through and through (please see enclosed pictures).
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: a device history record review was completed by our quality engineer team for provided material number 303367 and lot number 2021654.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.
|
| |
|