Catalog Number 301029
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/27/2023
Event Type  malfunction  
Event Description
It was reported that moisture was found in an unspecified amount of bd luer-lok¿ tip syringes.The following information was provided by the initial reporter: "we have a recent complaint from a customer that is seeing significant moisture inside all of syringes in their packs.".
 
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
Investigation summary: two photos were provided to our quality team for investigation.Upon examination of the returned photos, it was observed that silicone visible on the tip of the stopper.From the photos provided the amount of silicone is acceptable per product specification.Silicone is an inert, non-toxic medical substance used as a lubricant for disposable hypodermic products.It is an integral part of the syringe, enabling it to perform as required in various clinical applications and does not present a safety or efficacy issue nor does it impact product function.A device history record review was completed for provided lot number 2201726.A review showed no rejected inspections or quality issues during the production that could have contributed to the reported defect.
 
Event Description
It was reported that moisture was found in an unspecified amount of bd luer-lok¿ tip syringes.The following information was provided by the initial reporter: "we have a recent complaint from a customer that is seeing significant moisture inside all of syringes in their packs.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD LUER-LOK¿ TIP SYRINGE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16729046
MDR Text Key313306850
Report Number1213809-2023-00356
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K980987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number301029
Device Lot Number2201726
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2023
Initial Date FDA Received04/12/2023
Supplement Dates Manufacturer Received04/28/2023
Supplement Dates FDA Received05/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-