| Model Number |
SONATA |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The parents reported that the recipient had a blow to her head on (b)(6) 2023.After the accident, the recipient's hearing performance with the device was not good.The recipient has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the investigation results revealed overload fractures in the active electrode confirming that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.Other damages found during device investigation are most likely related to the explantation surgery.This is a final report.
|
| |
|
Event or Problem Description
|
|
The parents reported that the recipient had a blow to her head on 01-mar-2023.After the accident, the recipient_s hearing performance with the device was not good.The recipient has been re-implanted.
|
| |
|