Catalog Number 538421
Medical Device Problem Codes Crack (1135); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the manufacturing documentation associated with lot 18131652 presented no issues during the manufacturing process that can be related to the reported event.This device is available for analysis, but the engineering report is not yet available.However, it will be submitted within 30 days upon receipt.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, the tip of a 4f judkins right jr 4 100cm infiniti diagnostic catheter was found cracked during the procedure.There were no reports of patient injury.The device will be returned for evaluation.
 
Additional Manufacturer Narrative
As reported, the tip of a 4f judkin's right (jr) 4 100cm infiniti diagnostic catheter was found cracked during the procedure.There were no reports of patient injury.One non-sterile unit of diagnostic catheter (cath 4f inf 100cm jr4) was received for evaluation.During visual inspection, the hub was found to be broken.No other anomalies were observed.The fusion area of the catheter was observed under the vision system and no anomalies were found.Sem analysis was performed to determine the root cause of the separation to the luer hub.The analysis showed striation patterns along the luer hub proximal end.A product history record (phr) review of lot 18131652 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The complaint reported by the customer as ¿brite tip/distal tip-cracked¿ was not confirmed.No anomalies were observed at the fusion area or on the tip of the catheter.The event "luer hub-separated¿ was confirmed.A separation was found on the hub.The exact cause of the separation cannot be conclusively determined however, striation patterns are commonly associated with separations caused by material overload.Therefore, it is assumed that the material was induced to forces that exceeded the material yield strength prior to the separation.Storage and/or handling factors are likely causes of the event.Users are trained to inspect for signs of damage prior to and during use.Any product with damage is not to be used.Information for safety is provided in the products labeling with the intent to make the user aware of the risks.According to the instructions for use (ifu), although not intended as a mitigation of risk, ¿store in a cool, dark, dry place.Do not use if package is open or damaged.¿ based on the information available and the phr review, there is no indication that the event is related to the device design or manufacturing process.Therefore, no preventative or corrective actions will be taken at this time.
 
Additional Manufacturer Narrative
An updated review of the manufacturing documentation associated with lot 18131652 presented no issues during the manufacturing process that can be related to the reported event.This device has been analyzed but the final, approved draft of the engineering report and its conclusion is not yet available.However, it will be submitted within 30 days upon receipt.
 
Additional Manufacturer Narrative
-after further review of additional information received the following sections have been updated accordingly: d2, g1, g3, g6, h1, h2, h3, h6, and h10 -this device has been analyzed but the final, approved draft of the engineering report and its conclusion is not yet available due to need for additional testing.However, it will be submitted within 30 days upon receipt.-additional information is pending and will be submitted within 30 days upon receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes, florida 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes, florida 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, florida 33014
7863138372
MDR Report Key16731560
Report Number9616099-2023-06467
Device Sequence Number20543792
Product Code DQO
UDI-Device Identifier10705032014618
UDI-Public(01)10705032014618(17)250630(10)18131652
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K862244
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Catalogue Number538421
Device Lot Number18131652
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2023
Supplement Date Received by Manufacturer06/02/2023
08/01/2023
09/14/2023
Initial Report FDA Received Date04/13/2023
Supplement Report FDA Received Date06/27/2023
08/25/2023
09/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Patient SexUnknown
-
-