• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER Back to Search Results
Model Number 996101
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was not returned.
 
Event or Problem Description
It was reported that after opening the product, when the customer started to inflate the balloon, the doctor noticed the balloon was leaking.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned and further investigation was not conclusive.Though a specific cause cannot be identified, a potential root cause for this event could be, "poor balloon design".The device was used for treatment purposes.It is unknown if the device had met relevant specifications or resulted in the reported event.No manufacturing issues or non-conformances were noted during review of the dhr that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "if resistance is felt when removing either the catheter or the guidewire from the working sheath, stop and consider removing them as a single unit to prevent damage to the product.Applying excessive force to the catheter can result in tip breakage or balloon separation." "inspection prior to use the x-force® nephrostomy balloon dilation catheter is a sterile, single use device.Carefully inspect the catheter and the sterile packaging for signs of damage that may have occurred during shipment.Do not use the product if damage is evident." "storage store in a cool, dry and dark place.Do not store near radiation or ultraviolet light sources as these may damage product materials." "preparation of the catheter all x-force® nephrostomy balloon dilation catheters contain air in the balloon lumen.The air must be removed to allow liquid to fill the balloon when it is inflated.1.Remove the protective sheath from the balloon.2.Attach the inflation device to the connector on the balloon lumen.3.Open the stopcock, and draw back on the inflation device to remove the air from the balloon catheter 4.Close the stopcock, remove the inflation device, depress the plunger to remove any air and reattach to the balloon catheter.5.Repeat steps 3-4 until all air is removed from the balloon lumen." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that after opening the product, when the customer started to inflate the balloon, the doctor noticed the balloon was leaking.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER
Common Device Name
DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16732507
Report Number1018233-2023-02537
Device Sequence Number15119191
Product Code LJE
UDI-Device Identifier10801741127660
UDI-Public(01)10801741127660
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K063632
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/07/2024
Device Model Number996101
Device Catalogue Number996101
Device Lot NumberBMGWFM23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2023
Supplement Date Received by Manufacturer07/17/2023
Initial Report FDA Received Date04/13/2023
Supplement Report FDA Received Date07/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;