• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR W SIDNE (EU1); INSUFFLATOR, LAPAROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR W SIDNE (EU1); INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Catalog Number 0620040611
Device Problems Increase in Pressure (1491); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2023
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was overpressure.
 
Manufacturer Narrative
The device was received at the local service facility for investigation.The technical service report is attached (see comm log), and indicates: ro#: (b)(4).Account: genmed gloucester.Reported by the customer: ref: (b)(4).An error message halfway through surgery saying there was an occlusion with the gas.Surgeon was unable to find the issue with the gas tubing so assume it¿s a fault with the machine.It came up with a red symbol which stopped the gas and stopped us using it.Happened twice last week but resolved itself prior to a repair job being rasied p/n #: 0620040611.S/n: (b)(4).Summary of evaluation: device received out of calibration ¿ calibration on this device was due on 1/6/2021.Successfully completed calibration and qip.Device will be shipped back to the customer.Probable root cause: system design (tolerance stack up issue).Manufacturing / assembly/ service error.Use error.Damage during shipping (inadequate packaging).Damage during shipping (abuse during transit).Tubeset locking mechanism malfunction.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Event Description
It was reported that there was overpressure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR W SIDNE (EU1)
Type of Device
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16734763
MDR Text Key313294538
Report Number0002936485-2023-00309
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K063367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0620040611
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/22/2023
Initial Date FDA Received04/13/2023
Supplement Dates Manufacturer Received03/22/2023
Supplement Dates FDA Received11/10/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-