This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue, and ensure any required mdr reporting is completed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, the following are the potential causes: damage to the charge coupled device (ccd) unit, damage or deformation of the connector, movable l-range sliding error that occurred in the zoom scope, blurred image occurred within the allowable range.The event can be detected/prevented by following the instructions for use: operation manual; inspection of the endoscopic system.Operation manual; important information: please read before use warnings and cautions.During the device evaluation, service found the specified resolving power was not obtained, due to damage to the charge coupled device (ccd) unit.The objective lens was cracked, the adhesive on the bending cover (a-rubber) was worn, the electrical connector was damaged, the bending angle in the up direction did not meet specifications; due to wear of the angle wire.The up/down knob could not be locked securely; due to wear of the forceps elevator lever.Per the legal manufacturer, these other device issues identified by service have no potential to cause or contribute to death or serious injury if the malfunctions were to recur.Olympus will continue to monitor field performance for this device.
|
|
The subject device was received, at an olympus service center for evaluation, with a report that the image was greenish.There was no patient or user injury reported.During inspection and testing, service found the endoscopic image was blurry.This report is being submitted for the malfunction [blurred image] found during the device evaluation.
|