MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problems
Obstruction of Flow (2423); Infusion or Flow Problem (2964)
|
Patient Problems
Scar Tissue (2060); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Type
Injury
|
Event Description
|
Information was received from a healthcare provider (hcp) via a manufacturer representative (rep) regarding a patient receiving dilaudid (unknown dose and concentration) via an implantable pump.The indications for use were unknown.It was reported that there was a lack of efficacy.The environmental, external, or patient factors that may have led to or contributed to the issue were unknown.A dye study was performed, and they were unable to withdraw from the catheter access port (cap).The catheter and pump were replaced.The hcp noted that the issue was the catheter was scarred down and pinched off in multiple spots.That part was still in the patient and the old catheter was tied off.The pump was replaced just in case.It was noted that the hcp had no further information.The patient weight and medical history were asked but unknown.The patient status was alive with noinjury.The issue was resolved.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id:8780 lot# serial# (b)(4).Implanted: (b)(6) 2021.Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4)., ubd: 29-jun-2023, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: analysis of the catheter revealed no significant anomaly; miscellaneous acceptable testing-catheter incomplete/returned in segments.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|