• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH STR BALL SPIKE FOR LOW PRO PELVIC SYS; RETRACTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH STR BALL SPIKE FOR LOW PRO PELVIC SYS; RETRACTOR Back to Search Results
Model Number 03.100.018
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on an unknown date, that the wood handled instruments are leaving a stain/residue when they are peel-packed and sterilized.There were no patient consequence.This complaint involves fourteen (14) devices.This report is for one (1) str ball spike for low pro pelvic sys.This is report 5 of 14 for complaint (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional narrative: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is a j&j sales representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the str ball spike for low pro pelvic sys was worn from the handle and the shaft surface.The device show traces of usage, the handle is worn, and the shaft shows small scratched due usage.No traces of foreign substance were found.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.After a visual inspection per guidance provided in windchill document, it was determined that the reusable instrument device was worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the str ball spike for low pro pelvic sys would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Device: part number: 03.100.018.Lot number: t156350.Manufacturing site: (b)(4).Release to warehouse date: 17-oct-2017.A review of the device history records was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STR BALL SPIKE FOR LOW PRO PELVIC SYS
Common Device Name
RETRACTOR
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16738387
Report Number8030965-2023-04632
Device Sequence Number20548689
Product Code GAD
UDI-Device Identifier10886982070678
UDI-Public(01)10886982070678
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.100.018
Device Catalogue Number03.100.018
Device Lot NumberT156350
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2023
Supplement Date Received by Manufacturer05/18/2023
Initial Report FDA Received Date04/13/2023
Supplement Report FDA Received Date05/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/17/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4.0MM HEXAGONAL SCREWDRIVER; CANNULATED 4.0MM HEXAGONAL SCREWDRIVER; DHS/DCS COUPLING SCREW; DHS/DCS COUPLING SCREW; GOUGE W/10 L205/90; LARGE HEXAGONAL SCREWDRIVER; LARGE HEXAGONAL SCREWDRIVER; LARGE HEXAGONAL SCREWDRIVER LONG; SCRDRIVER-HEX-SMALL W/HOLD-SLEEVE; SCRDRIVER-HEX-SMALL W/HOLD-SLEEVE; SCRDRIVER-HEX-SMALL W/HOLD-SLEEVE; SCRDRIVER-HEX-SMALL W/HOLD-SLEEVE; SDDRIVE SCREWDRIVER T8; SDDRIVE SCREWDRIVER T8
Patient SexUnknown
-
-