The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the operation channel (primary) stuck accessory/object.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the operation channel (primary).In addition, our technician confirmed that the insertion flexible tube coating damage, the lcb distal cover glass cracked, the bending rubber cut, and the angulation down angulation failure; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|