PRO MED INSTRUMENTS GMBH DORO® SKULL CLAMP RADIOLUCENT Back to Search Results
Model Number 3034-00
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code Laceration(s) (1946)
Type of Reportable Event Serious Injury
Event or Problem Description
Distributor informed our service department on the 24th of march that one of our products was involved in a case where a laceration occured.
 
Additional Manufacturer Narrative
As the device was in specification and did not show any deviation that could cause the reported incident, we suspect, that maybe the pinning technique has been not optimal as described in the instruction manual: "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline." the interval of the supplier maintenance was exceeded by one year by the customer.Although no deviations were found in this case, it cannot be excluded in general that any deviations may remain hidden if this maintenance specification is not complied with, and that possible risks may therefore arise.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® SKULL CLAMP RADIOLUCENT
Common Device Name
DORO® SKULL CLAMP RADIOLUCENT
Manufacturer (Section D)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
nicholas preissler
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
MDR Report Key16741687
Report Number3003923584-2023-00014
Device Sequence Number17802684
Product Code HBL
UDI-Device Identifier04250435502181
UDI-Public04250435502181
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K063494
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number3034-00
Device Catalogue Number3034-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2023
Initial Report FDA Received Date04/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/29/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-